Compact silicon biosensor for the clinical range estimation of blood serum triglycéride
, Mohanasundaram Sulur Veeramani, Noel Prashant Ratchagar, ,
Published in IEEE Computer Society
2013
Abstract
Electrolyte Insulator Semiconductor Capacitor (El S CAP) is a transducer that can detect triglycérides (TGs) in blood serum through an enzymatic response that changes the pH of the electrolyte (blood serum). This change in pH is transformed by the EISCAP to a shift in the capacitance-voltage (C-V) characteristics of the device which in turn can be mapped to the amount of TG in blood serum. We have developed a miniaturized biochip with immobilized enzyme integrated with a thin film counter electrode for the estimation of TG in blood serum within the clinical range. The miniaturized devices are tested using blood serum sample to estimate the TG concentration and compared with clinical data. © 2013 IEEE.