Use of magneto plethysmogram sensor for real-time estimation of hemoglobin concentration
Published in Institute of Electrical and Electronics Engineers Inc.
2021
Volume: 21
   
Issue: 4
Pages: 4405 - 4411
Abstract
A novel magnetoplethysmogram (MPG) system is designed and developed for the real time, noninvasive monitoring of hemoglobin concentration in arterial blood. The basic principle of modulated magnetic signature of blood (MMSB), is validated with clinical trial. The MPG probe consists of a permanent magnet and a Giant Magneto Resistance (GMR) sensor. A strap and stand design enable reliable capture of continuous cycles of MPG with relative ease. Clinical validation with data from 64 patients carried out at the institute hospital of IIT Madras, using the proposed system gave a worst-case error of 6.5% in the measurement of hemoglobin concentration in arterial blood. Linear regression analysis using least square method, exhibited a standard deviation of 0.64 g/dL. Results from the clinical validation establish the practicability of the proffered MPG system. © 2001-2012 IEEE.
About the journal
JournalData powered by TypesetIEEE Sensors Journal
PublisherData powered by TypesetInstitute of Electrical and Electronics Engineers Inc.
ISSN1530437X
Open AccessNo