Readout circuit design for an EISCAP biosensor
Published in Institute of Electrical and Electronics Engineers Inc.
2008
Pages: 73 - 76
Abstract
We present the design of readout electronics for an Electrolyte Insulator Semiconductor capacitor (EISCAP) sensor. The sensor detects the presence of bioanalytes by sensing the change in the pH of the electrolyte. Our technique uses an EISCAP relaxation oscillator embedded in a successive approximation analog-to-digital conversion algorithm to give a digital readout of the pH of the test solution. A microcontroller based pH readout instrument, incorporating calibration to account for sensor variations, is described. Experimental results that demonstrate the efficacy of our techniques are given. © 2008 IEEE.
About the journal
JournalData powered by Typeset2008 IEEE-BIOCAS Biomedical Circuits and Systems Conference, BIOCAS 2008
PublisherData powered by TypesetInstitute of Electrical and Electronics Engineers Inc.
Open AccessNo
Concepts (17)
  •  related image
    ANALOG-TO-DIGITAL CONVERSIONS
  •  related image
    BIOANALYTES
  •  related image
    DIGITAL READOUTS
  •  related image
    ELECTROLYTE INSULATOR SEMICONDUCTORS
  •  related image
    Micro controllers
  •  related image
    READ-OUT CIRCUITS
  •  related image
    READ-OUT ELECTRONICS
  •  related image
    Successive approximations
  •  related image
    TEST SOLUTIONS
  •  related image
    Analog to digital conversion
  •  related image
    Approximation algorithms
  •  related image
    Biosensors
  •  related image
    Electrolytes
  •  related image
    Readout systems
  •  related image
    RELAXATION OSCILLATORS
  •  related image
    Sensors
  •  related image
    Integrated circuit manufacture