Evaluation of fretting corrosion behaviour of CP-Ti for orthopaedic implant applications
Satendra Anand Kumar, , Suresh Krishnamoorthy Seshadri
Published in
2010
Volume: 43
   
Issue: 7
Pages: 1245 - 1252
Abstract
The fretting corrosion behaviour of CP-Ti in Ringer's solution was studied as a function of normal load, frequency and number of fretting cycles. The restoration ability of CP-Ti after the passive film is damaged due to fretting and as a function of the on-time/off-time ratio (intermittent fretting) was also evaluated. The change in free corrosion potential measured before the onset of fretting, with the onset of fretting, during fretting and after the fretting motion ceases, as a function of time, was used to evaluate the fretting corrosion behaviour. The restoration ability of CP-Ti after the passive film is damaged was ascertained by performing regression analysis of the potential data measured during repassivation. The morphological features of the fretted zone were assessed using scanning electron microscopy. Energy dispersive X-ray analysis was performed at the centre and edge regions of the fretted zone to identify their chemical nature. The study reveals that the excellent corrosion resistance and biocompatibility of CP-Ti are nullified under fretting conditions. Once the passive oxide layer is damaged due to fretting, repassivation is not instantaneous. The significant time delay in reaching the steady state potential implies that CP-Ti remains active and susceptible for corrosion. The difficulty in the instantaneous formation of passive film after the fretting induced damage of the passive film, dissolution of bare Ti from the damaged areas and the possible accumulation of the debris generated during fretting in the surrounding tissues raises concern on the safer use of CP-Ti as an implant material. © 2009 Elsevier Ltd. All rights reserved.
About the journal
JournalTribology International
ISSN0301679X
Open AccessNo
Concepts (33)
  •  related image
    Chemical nature
  •  related image
    Corrosion potentials
  •  related image
    CP-TI
  •  related image
    DAMAGED AREA
  •  related image
    EDGE REGION
  •  related image
    EXCELLENT CORROSION RESISTANCES
  •  related image
    FRETTING CYCLES
  •  related image
    FRETTING MOTION
  •  related image
    Function of time
  •  related image
    IMPLANT APPLICATION
  •  related image
    IMPLANT MATERIALS
  •  related image
    INDUCED DAMAGE
  •  related image
    Morphological features
  •  related image
    NORMAL LOADS
  •  related image
    Orthopaedic implants
  •  related image
    Passive films
  •  related image
    PASSIVE OXIDES
  •  related image
    RE-PASSIVATION
  •  related image
    RINGER'S SOLUTION
  •  related image
    STEADY-STATE POTENTIALS
  •  related image
    TIME RATIO
  •  related image
    Biocompatibility
  •  related image
    Corrosion resistance
  •  related image
    Corrosive effects
  •  related image
    Debris
  •  related image
    Dissolution
  •  related image
    FRETTING CORROSION
  •  related image
    Oxidation
  •  related image
    Passivation
  •  related image
    Regression analysis
  •  related image
    Restoration
  •  related image
    Scanning electron microscopy
  •  related image
    TITANIUM METALLOGRAPHY